Precision™ Spinal Cord Stimulator System Clinician Manual
Selection #2 – Impedances
When the "Impedances" option is selected from the Clinician Options menu, the Remote
Control will measure impedances at each contact on the lead(s). This will take about
10 seconds.
Note: During this measurement the serial number of the Stimulator will be displayed on the
screen.
The Remote Control will display the contact impedance status screen.
Contacts 1 through 8 (lead position "1-L") are represented by the rectangles in the top row,
starting with contact 1 on the left. Contacts 9 through 16 (lead position "2-R") are represented
by the rectangles in the bottom row, starting with contact 9 on the left.
Contacts within the typical impedance range are displayed as solid rectangles. High impedance
contacts (above 4500 ohms) are represented by hollow rectangles.