Removal of Trial Leads, Extensions, and Splitters

Remove bandages and properly cleanse the exit site. The method of removal depends upon
whether a temporary trial or permanent trial was performed.

Option A: Percutaneous Lead Removal after Temporary Trial

1. Clip sutures if used to secure the trial lead(s) in place.
2. Remove the lead(s) and discard.
3. To replace the trial lead(s) with permanent percutaneous lead(s), proceed to the instructions
for "Percutaneous Lead Placement in the Epidural Space" on page 17 or "Infinion CX
Lead Placement in the Epidural Space with Entrada Needle" on page 19.
4. To replace the trial lead(s) with a paddle lead proceed to the instructions for "Surgical
Paddle Lead Placement in the Epidural Space" on page 27.

Option B: Lead Extension Removal after Permanent Trial

1. Open the midline incision to expose the lead extension and connector.
2. Cut the lead extension at the connector. Do not cut the implanted lead.
Removal of Trial Leads, Extensions, and Splitters
Clinician Manual
91173656-02