Restoring Programs:

The original clinic programs can be restored by following the instructions below.
P
1. Press and hold
for approximately 3 seconds to access the Restore screen. Program 1
will be highlighted.
P
2. Press the
button to cycle through the programs and select the program to be restored.
3. Press p to restore the clinic-programmed settings.
Coverage Area Selection
Each program can consist of up to 4 different coverage areas (Areas). The stimulation strength
and other settings for each Area can be adjusted.
1. Press the
button to access the Area level screen. This controls the stimulation strength
"
for each Area individually. Press the
coverage area.
2. Press the p or q button to adjust the stimulation strength (amplitude) of the selected
Area.
Additional Area Options
The Pulse Width and Rate parameters are available as programming options to allow patients
greater control of their therapy. By default, access to Pulse Width and Rate are disabled in
the patient Remote Control. Access to these parameters can be enabled through the Bionic
Navigator software.
Rechargeable Stimulator System
button as needed to highlight a specific stimulation
"
Clinician Manual
91173656-02
59 of 385