2. If the Remote Control is not linked to a Stimulator, it will display a "Link?" screen. Press
link. The Remote Control will search for the nearest Stimulator.
If the Remote Control is already linked to a Stimulator, it may continuously search for it.
This link must be cleared before a new link can be established (See "Clinician Options" on
page 62).
3. The Remote Control will identify the Stimulator by serial number. Verify before proceeding.
P
4. Press
to confirm and continue. Press
When linking with a Trial Stimulator, the Remote Control will ask whether existing programs on
the Trial Stimulator should be cleared. If the programs on the Trial Stimulator are cleared and if
the Remote Control contains existing programs, it will then ask if these should be used.
When linking with an IPG, the Remote Control will automatically use any programs stored in the
IPG. If the IPG contains no programs and the Remote Control contains existing programs, it will
ask whether to use the programs from the Remote Control.
When the link is established, you will see the main Level screen:
Rechargeable Stimulator System
to cancel.
"
P
Clinician Manual
91173656-02
61 of 385
to