Precision™ Spinal Cord Stimulator System Clinician Manual
Securing the Trial Lead
Note: If using an Infinion CX Trial Lead, proceed to "Securing the Infinion CX Trial Lead" on
page 33.
1. Carefully withdraw the insertion needle from the epidural space by slowly pulling the needle
up towards the proximal end of the lead while holding the lead in place.
2. Once the insertion needle tip is exposed, hold the lead as close to the percutaneous exit
site as possible, then carefully pull the needle completely from the lead.
3. If desired, a suture may be used to close the wound and stabilize the lead.
4. Place and tape a stress relief loop and dress the wound.
5. If using a splitter, continue to "Lead Connection to Splitter" on page 24.
6. Continue with "Connecting to the Trial Stimulator" on page 41.