Precision™ Spinal Cord Stimulator System Clinician Manual
Permanent Lead Anchoring and Tunneling
Removing the Insertion Needle
Note: If using an Entrada Needle, proceed to "Removing the Entrada Needle".
1. Cut down around the insertion needle to provide access for anchoring the lead.
2. Carefully withdraw the insertion needle from the epidural space by slowly pulling the needle
up towards the proximal end of the lead while holding the lead in place.
3. Once the insertion needle tip is exposed, hold the lead as close to the exit site as possible,
then carefully pull the needle completely from the lead.
Removing the Entrada Needle
1. Carefully withdraw the slotted needle by slowly pulling the needle towards the proximal end
of the lead while stabilizing lead in line with the slot to keep lead in place.
Note: Do not reinsert the slotted needle into the sheath while the sheath is within the
body.
Stabilize lead in line with the slot.
Clinician Manual
91173656-02
34 of 385
15a
Slowly remove slotted needle.