Infinion CX Lead Placement in the Epidural Space with Entrada
Needle
Infinion CX Lead
The Infinion CX Lead has two tails to enable insertion into 8-contact IPG ports without the use
of the Splitter 2x8. The two tails will not fit through a standard 14 gauge insertion needle or
Epimed needle, therefore these needles are not compatible with the Infinion CX Lead. Do not
use a standard 14 gauge introducer or Epimed needle to introduce the Infinion CX Lead.
Use only the Entrada Needle with peelable sheath to introduce the Infinion CX Lead into the
epidural space.
An anchor must be pre-loaded on the distal end of the Infinion CX Lead prior to inserting the
lead into the Entrada Needle. When using an anchor or suture sleeve with the Infinion CX
Lead, the stylet may be in place within the lead when the anchor is loaded on the lead. Ensure
that stylet is removed from the lead prior to securing the anchor.
Entrada Needle
The Entrada Needle with peelable sheath is required to insert the Infinion CX Lead into the
epidural space. The Entrada Needle may also be used with all percutaneous leads listed under
"Product Model Numbers" on page ii.
Fully Assembled Entrada Needle
Components
Components are marked with dots to indicate use order, and the dots should be in the same
direction when assembling the needle.
Percutaneous Lead Placement in the Epidural Space
Needle Stylet, •
Loss of Resistance (LOR) adaptor, ••
Slotted Needle, •••
Peelable Sheath
Clinician Manual
91173656-02
19 of 385