Tunneling The Lead Or Lead Extension - Boston Scientific Precision SC-1110 Mode D'emploi

Système de stimulation médullaire
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Precision™ Spinal Cord Stimulator System Clinician Manual
Note: The 4 cm and 2.3 cm suture sleeves each have three (3) compression slots, which are
designed to reduce slippage.
4. Suture the sleeve to the supraspinous ligament or deep fascia through the suture sleeve
holes.
5. Tie multiple sutures as tightly as possible around the suture sleeve to secure it to the lead.
CAUTION: Tightening sutures directly on the lead can damage the lead.
6. For Permanent Trials, proceed to the instructions for "Tunneling the Lead or Lead
Extension" on page 36.
7. For Permanent IPG Implantation, proceed to the instructions for "IPG Implantation" on
page 45.

Tunneling the Lead or Lead Extension

Note: If tunneling Infinion CX Lead(s), it is recommended to use the Long Tunneling Tool
(35 cm).
1. Attach the tunneling tool handle to the shaft by turning the locking mechanism clockwise.
Tool Handle
Locking Mechanism
Clinician Manual
91173656-02
36 of 385
Shaft