For Surgical Paddle Lead Connection

• Left side of lead connects to IPG port "1-L".
• Right side of lead connects to IPG port "2-R".
Note: The right tail of the Paddle Lead is laser-etched to allow for easy identification of right
and left.
Left
Right
Upper
1L
2R
1L
1. Fully insert the lead(s), extension(s), or splitter(s) into the IPG port(s), being careful not
to stress or bend the proximal end of the lead or extension. When the lead is properly
inserted, the lead will stop and the retention ring will be located under the setscrew.
Note: If you experience difficulty when inserting the lead, lead extension, or port plug, use the
hex wrench to loosen (counterclockwise) the setscrew and/or gently rotate the lead to
help advance the proximal end.
Note: To confirm good connections, check impedances before tightening the setscrew.
Lower
Single
2R
1L
Connecting to the IPG
Clinician Manual
91173656-02
47 of 385