3. Remove the extension, being careful not to contact non-sterile portions to the patient's
body.
4. Loosen the extension connector setscrew using the hex wrench. Disconnect and remove
the connector without moving the implanted lead.
5. Proceed to the instructions for "IPG Implantation" on page 45.

Option C: Splitter Removal after Trial

1. If the splitter has been implanted, open the incision to expose the lead and splitter junction.
2. Loosen the connector set screw on the splitter receptacles using the hex wrench provided.
3. Disconnect and discard the splitter components.