10. If using the Splitter 2x8, prior to closing the wound, clean the top of the splitter set-screw
seal plug and use silicone medical adhesive (e.g. Dow Corning Silastic
Silicone, Type A - Sterile, as available from Boston Scientific, part number SC-4320) to coat
and seal the top of the seal plug that has been penetrated by the hex wrench.
Note: Inadvertent damage to the septum seal may lead to unintended stimulation at the
splitter 2x8 if the Medical Adhesive is not used as intended.
11. Proceed to "Connecting to the Trial Stimulator" on page 41 or "Connecting to the IPG" on
page 46.