Precision™ Spinal Cord Stimulator System Clinician Manual
6. OPTIONAL. If exchange of the lead stylet is desired, carefully pull out the existing stylet
and insert the preferred stylet. While inserting the stylet into the lead, if resistance is
encountered, withdraw the stylet approximately 3 cm, rotate the lead and/or stylet, and
gently advance the stylet. If resistance is still encountered, repeat the above procedure until
the stylet can be fully inserted.
WARNING: Do not exchange the lead stylet while the electrode array of the lead is in the
bevel of the insertion needle. If the electrode array is in the bevel area, remove the lead
from the insertion needle before exchanging the stylet. Inserting the lead stylet in the lead
while the electrode array is in the bevel of the insertion needle increases the risk of lead
and tissue damage.
WARNING: If the lead stylet is removed and reinserted, do not use excessive force when
inserting the stylet into the lead. The use of instruments, such as forceps, to grasp the stylet
during insertion is not recommended as this could result in applying excessive force and
could increase the risk of lead and tissue damage.
7. Advance the lead to the appropriate vertebral level under fluoroscopic guidance. A sufficient
length of lead (i.e., at least 10 cm, or approximately three vertebrae) should reside in the
epidural space to aid in lead stabilization.
8. If use of a splitter is desired or you are using the Infinion™ 16 lead, continue "Lead
Connection to Splitter" on page 24. Otherwise, continue to "Connecting the OR Cable
Assembly" on page 28.
Clinician Manual
91173656-02
18 of 385