T T r r a a n n s s f f e e r r r r i i n n g g o o f f p p a a t t i i e e n n t t f f r r o o m m o o n n e e p p a a t t i i e e n n t t s s u u p p p p o o r r t t p p l l a a t t f f o o r r m m t t o o a a n n o o t t h h e e r r
Do not exceed the safe working load of the product.
C C A A U U T T I I O O N N
• Do not use the support surface as a transfer device to avoid the risk of patient injury.
• Do not stick needles into a support surface through the support surface cover. Holes may allow body fluids to enter the
inside (inner core) of the support surface and could cause cross-contamination, product damage, or product
malfunction.
P P r r e e r r e e q q u u i i s s i i t t e e : : Follow hospital protocols required to transfer a patient from one surface to another.
To transfer the patient from one patient support surface to another:
1. Position one patient support platform next to the other patient support platform. Minimize the space between the two
platforms.
2. Apply the brakes for both patient support platforms.
3. Inflate the support surface with max inflate, recommended. See the P P r r o o C C u u i i t t y y Operations Manual to start and stop max
inflate.
4. Adjust the patient support platform heights so that they are level with one another.
5. Transfer the patient. Follow hospital protocols for applicable safety rules for patient and operator.
6. Stop max inflate.
S S t t o o r r i i n n g g t t h h e e p p r r o o d d u u c c t t
Before you store the product, deflate the support surface by activating CPR.
Surface de soutien. destinée à une utilisation avec le châssis de lit intouch fl27 (2131, 2141, 2151, 2152, version logicielle 4.0 ou supérieure). (72 pages)