Télécharger Imprimer la page

Stryker Isolibrium PE Manuel D'utilisation page 21

Surface de support

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 267
T T r r a a n n s s f f e e r r r r i i n n g g o o f f p p a a t t i i e e n n t t f f r r o o m m o o n n e e p p a a t t i i e e n n t t s s u u p p p p o o r r t t p p l l a a t t f f o o r r m m t t o o a a n n o o t t h h e e r r
Do not exceed the safe working load of the product.
C C A A U U T T I I O O N N
• Do not use the support surface as a transfer device to avoid the risk of patient injury.
• Do not stick needles into a support surface through the support surface cover. Holes may allow body fluids to enter the
inside (inner core) of the support surface and could cause cross-contamination, product damage, or product
malfunction.
P P r r e e r r e e q q u u i i s s i i t t e e : : Follow hospital protocols required to transfer a patient from one surface to another.
To transfer the patient from one patient support surface to another:
1. Position one patient support platform next to the other patient support platform. Minimize the space between the two
platforms.
2. Apply the brakes for both patient support platforms.
3. Inflate the support surface with max inflate, recommended. See the P P r r o o C C u u i i t t y y Operations Manual to start and stop max
inflate.
4. Adjust the patient support platform heights so that they are level with one another.
5. Transfer the patient. Follow hospital protocols for applicable safety rules for patient and operator.
6. Stop max inflate.
S S t t o o r r i i n n g g t t h h e e p p r r o o d d u u c c t t
Before you store the product, deflate the support surface by activating CPR.
2973-009-005 Rev AF.0
17
EN

Publicité

loading

Ce manuel est également adapté pour:

297300000000