• Do not use the support surface as a transfer device to avoid the risk of patient injury.
• Always unplug the support surface power/data cable before you clean or disinfect.
• Do not immerse the support surface or foot box in cleaning or disinfectant solutions.
• Do not allow fluid to pool on the support surface. Fluids may cause degradation of components and may cause
unpredictable safety and performance of this product.
• Always inspect the support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned
zippers as you clean the covers. If the cover is compromised, remove the support surface from service immediately and
replace the cover to prevent cross-contamination.
• Always perform preventive maintenance based on the use of the product. An increase in the use of the product may
include more frequent cleaning and disinfection and may adversely affect the life of the support surface.
• Always wipe each product with clean water and dry after you clean or disinfect. Some cleaning and disinfecting agents
are corrosive in nature and may cause damage to the product. Failure to properly rinse and dry the product leaves a
corrosive residue on the surface of the product and may cause premature degradation of critical components. Failure to
follow these cleaning or disinfecting instructions may void your warranty.
• Do not iron, dry-clean, tumble dry, or power wash the support surface or support surface covers.
• Always disinfect the support surface between patients, to avoid the risk of cross-contamination and infection.
• Do not overexpose the covers to higher concentration disinfectant solutions as these may degrade the covers.
• Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers greater than 3% glycol ether as
they may damage the cover or the CPR strap print.
• Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the product. If such
use is necessary, carefully observe stacked or adjacent equipment to make sure that they operate properly.
• The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
2973-009-005 Rev AF.0