Télécharger Imprimer la page

Stryker Isolibrium PE Manuel D'utilisation page 30

Surface de support

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 267
E E M M C C i i n n f f o o r r m m a a t t i i o o n n
C C A A U U T T I I O O N N
• Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the product. If such
use is necessary, carefully observe stacked or adjacent equipment to make sure that they operate properly.
• The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
N N o o t t e e - - Portable RF communications equipment, including peripherals such as antenna cables and external antennas,
should be no closer than 12 inches (30 cm) to any part of I I s s o o l l i i b b r r i i u u m m PE support surface, including cables specified by the
manufacturer.
The Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface was evaluated using the following cables:
Isolibrium to bed
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ' ' s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c e e m m i i s s s s i i o o n n s s
The Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface is intended for use in the electromagnetic environment
specified below. The customer or the user of the Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface should assure that
it is used in such an environment.
E E m m i i s s s s i i o o n n s s t t e e s s t t
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
The Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface is suitable for use in a professional healthcare facility
environment and not in environments exceeding immunity test conditions that the product was evaluated to, such as near
high frequency (HF) surgical equipment and inside of the radio frequency (RF) shielded room of magnetic resonance
imaging (MRI) equipment. The customer or the user of the Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface should
assure that it is used in such an environment and that the electromagnetic environment guidance listed below is followed.
I I m m m m u u n n i i t t y y t t e e s s t t
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrostatic fast transient/
burst
IEC 61000-4-4
EN
26
C C a a b b l l e e
C C o o m m p p l l i i a a n n c c e e
Group 1
Class A
Class A
Complies
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ' ' s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y
I I E E C C 6 6 0 0 6 6 0 0 1 1 t t e e s s t t l l e e v v e e l l
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
L L e e n n g g t t h h ( ( m m ) )
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c e e n n v v i i r r o o n n m m e e n n t t
N N o o t t e e - - The emissions characteristics of this
equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B
is normally required) this equipment might not offer
adequate protection to radio-frequency
communication services. The user might need to take
mitigation measures, such as relocating or re-
orienting the equipment.
C C o o m m p p l l i i a a n n c c e e l l e e v v e e l l
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
1.0
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c
e e n n v v i i r r o o n n m m e e n n t t - - g g u u i i d d a a n n c c e e
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Main power quality should
be that of a typical
commercial or hospital
environment.
2973-009-005 Rev AF.0

Publicité

loading

Ce manuel est également adapté pour:

297300000000