Wavewriter Alpha And Wavewriter Alpha Prime Maximum Current Amplitude Per Electrode Vs Impedance; For The 4X8 Surgical Lead - Boston Scientific Bionic Navigator 3D 4.0 Manuel De Programmation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 82
Bionic Navigator™ 3D 4.0 Programming Manual
WaveWriter Alpha and WaveWriter Alpha Prime Maximum Current Amplitude per
Electrode vs Impedance
For all Leads other than the 4x8 Surgical Leads
Maximum Amplitude Based on Impedance and Pulse Width
Programmable up to 25.5 mA
25.50
Programmable up to 21.1 mA
20.40
Programmable up to 15.8 mA
15.30
Programmable up to 12.7 mA
10.20
5.10
0.00
300
400

For the 4x8 Surgical Lead

Bionic Navigator™ 3D 4.0 Programming Manual
92395591-02 26 of 1595
500
600
700
Impedance (Ω)
800
900
1000
1100
Pulse Width:
PW = 20 us
20 µsec
PW = 100 us
100 µsec
PW = 200 us
200 µsec
PW = 300 us
300 µsec
PW = 400 us
400 µsec
PW = 600 us
600 µsec
800 µsec
PW = 800 us
PW = 1000 us
1000 µsec