Connecting to the IPG

For Percutaneous Lead - Dual
Lead Connection
• Superior (upper or left) leads, extensions,
or 2x4 splitters connect to IPG port "1-L".
• Inferior (lower or right) leads, extensions,
or 2x4 splitters connect to IPG port "2-R".
For Infinion™ Lead Connection
• Laser-marked tail of Splitter 2x8 connects
to IPG port "1-L".
• Unmarked tail of Splitter 2x8 connects to
IPG port "2-R"
For Percutaneous Lead - Single
Lead Connection
• Connect a single lead, extension, or 2x4
splitter to IPG port "1-L".
• Insert port plug into unused port.
Connecting to the IPG
For Surgical Paddle Lead
Connection
• Left side of lead connects to IPG port
"1-L".
• Right side of lead connects to IPG port
"2-R".
Note: The right tail of the Paddle Lead
is laser-etched to allow for easy
identification of right and left.
Left
Right
Upper
1L
2R
1L
1. Fully insert the lead(s), extension(s), or
splitter(s) into the IPG port(s), being care-
ful not to stress or bend the proximal end
of the lead or extension. When the lead is
properly inserted, the lead will stop and
the retention ring will be located under the
setscrew.
90607862-22 Rev A
Lower
Single
2R
1L
Clinician Manual