Implantable Pulse Generator
Connecting the Lead, Extension, Splitter, or Connector to the IPG
IPG ports are labeled as follows:
For convenience, connect leads or splitter tails to the IPG ports corresponding to their locations, superior
versus inferior or left versus right lead placements. For example:
• Superior leads to upper IPG ports A or B. Inferior leads to IPG ports C or D.
• For the Infinion™ 16, connect the splitter tail with laser etched bands (contacts 1-8 of the Infinion
16 lead) to the left ports A or C and the unmarked splitter tail (contacts 9-16 of the Infinion 16 lead)
to the right ports B or D.
• For the Infinion CX Lead, connect the tail with the single marker band (contacts 1-8 of the Infinion
CX Lead) to the left ports A or C and the tail with two marker bands (contacts 9-16 of the Infinion
CX Lead) to the right ports B or D.
• For the Artisan 2x8 surgical lead, connect the left side to the left ports A or C. Connect the right side
(the laser-etched tail), contacts 9-16, to the right ports B or D.
• For the 4x8 Surgical Lead, connect the lead tail as follows:
- Lead tail with one marker band (contacts 1-8) to port A.
- Lead tail with two marker bands (contacts 9-16) to port B.
- Lead tail with three marker bands (contacts 17-24) to port C.
- Lead tail with four marker bands (contacts 25-32) to port D.
Example: 4x8 Surgical Lead (electrodes faced down)
4
3
2
1
1
Contacts 1-8
2
Contacts 9-16
3
Contacts 17-24
4
Contacts 25-32
Implantable Pulse Generator
92177206-07
8 of 53
5
Connects to port A
6
Connects to port B
7
Connects to port C
8
Connects to port D
8
7
6
5