Connecting the Lead, Extension, Splitter, or Connector to the IPG

Note: Lead extensions, adapters, and splitters are not full-body MRI scan eligible.
IPG ports are labeled as follows:

Connecting to the IPG:

• Leads connect to IPG ports C or D.
• For the Infinion 16, connect the splitter tail with laser-etched bands (contacts 1-8 of the Infinion
16 lead) to the left port C and the unmarked splitter tail (contacts 9-16 of the Infinion 16 lead) to
the right port D.
• For the Infinion CX Lead, connect the tail with the single marker band (contacts 1-8 of the Infinion
CX Lead) to the left port C and the tail with two marker bands (contacts 9-16 of the Infinion CX
Lead) to the right port D.
• For the Artisan 2x8 surgical lead, connect the left side to the left port C. Connect the right side
(the laser-etched tail), contacts 9-16, to the right port D.
1. Fully insert the lead(s), extension(s), splitter(s), and/or connector(s) into the IPG port(s), being careful
not to stress or bend the proximal end of the lead. When the lead is properly inserted, the lead will stop
and the retention ring will be located under the set screw.
Note: Lead extensions, adapters, and splitters are not full-body MRI scan eligible
2. Fully insert a port plug into unused IPG ports.
Note: If you experience difficulty when inserting the lead, lead extension, splitter, connector, or port plug,
use the hex wrench to loosen (counterclockwise) the set screw and/or gently rotate the lead to help
advance the proximal end.
Note: To confirm good connections, go to the Systems Settings menu in the FreeLink Remote Control
and perform a Lead Check. Ensure that all boxes in the Lead Check screen are checked GREEN
before tightening the set screws. If one or more RED X appears the patient may not be eligible for
a full body MRI scan. The IPG must be in contact with the subcutaneous pocket in order to receive
accurate impedance measurements.
Production
Boston Scientific Confidential. Unauthorized use is prohibited.
Montage MRI IPG Entrada 2.0 19L
Instructions for Use
Implantable Pulse Generator
91053242-08 Rev A