Percutaneous Leads Directions for Use

Lead/Extension/Splitter Removal

Remove bandages and properly cleanse the exit site. The method of removal depends upon whether
a temporary trial or permanent trial was performed.
Option A: Percutaneous Lead Removal after Temporary Trial
1. Clip sutures if used to secure the trial lead(s) in place.
2. Remove the lead(s) and discard.
3. To replace the trial lead(s) with permanent percutaneous lead(s), proceed to the instructions for
"Percutaneous Lead Placement in the Epidural Space" on page 7.
4. To replace the trial lead(s) with a paddle lead proceed to instructions for "Surgical Paddle Lead
Placement in the Epidural Space" in the Surgical Leads DFU.
Option B: Lead Extension Removal After Permanent Trial
1. Open the midline incision to expose the lead extension and connector.
2. Cut the lead extension at the connector. Do not cut the implanted lead.
3. Remove the extension, being careful not to contact non-sterile portions to the patient's body.
4. Loosen the extension connector setscrew using the hex wrench. Disconnect and remove the connector
without moving the implanted lead.
5. Proceed to the IPG Implantation instructions as follows:
SCS System
Precision SCS System
Precision SCS System with
MultiWave Technology
Precision Spectra SCS System
Precision Novi SCS System
Percutaneous Leads Directions for Use
90970887-03 Rev A
18 of 441
Reference Document for instructions
Precision Spinal Cord Stimulator Clinician Manual
Precision Spinal Cord Stimulator System with MultiWave
Technology Clinician Manual
Precision Spectra IPG DFU
Precision Novi IPG DFU
Section Title
"IPG Implantation"
"IPG Implantation"
"IPG Implantation"
"IPG Implantation"