Tunneling The Lead Or Lead Extension - Boston Scientific Precision Manuel

Masquer les pouces Voir aussi pour Precision:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Tunneling the Lead or Lead
Extension
1. Attach the tunneling tool handle to the
shaft by turning the locking mechanism
clockwise.
Tool Handle
Locking Mechanism
2. Mark the desired route of the tunnel.
3. Administer the appropriate local anesthetic
along the tunneling path.
Permanent Lead Anchoring and Tunneling
4. OPTIONAL. If necessary, bend the tool
shaft to conform to the patient's body.
5. Make a small incision at the desired exit
site.
6. Create a subcutaneous tunnel between
the midline incision and the exit site until
the straw is visible and accessible at the
exit point.
Shaft
7. Unscrew and remove the tunneling tool
handle.
Clinician Manual
90607862-22 Rev A
29 of 303