Preparing The Bair Hugger Model 875 Patient Adjustable Warming Unit For Use; Placing The Patient Adjustable Warming Unit On The Wall-Mount Bracket; Mounting The Patient Adjustable Warming Unit On An Iv Pole; Mounting The Patient Adjustable Warming Unit On A Bedrail - 3M Bair Hugger 875 Manuel De L'utilisateur

L'unité de réchauffement réglable par le patient
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
Do not modify this equipment without authorization from the
manufacturer.
Over torqueing the suction wall mount adapter may cause
device damage.
Preparing the Bair Hugger Model 875 Patient
Adjustable Warming Unit for Use
Before you begin using the patient adjustable warming unit, make
sure that it is safely placed on a flat, hard surface such as a table, or
securely mounted on a wall, IV pole, suction wall mount adapter, or
bedrail.
Caution: Do not place the patient adjustable warming unit on a
soft uneven surface, such as a bed, or a visibly wet surface as the
air intake may become blocked and cause the patient adjustable
warming unit to overheat, compromising the patient adjustable
warming unit's performance.
A wall-mount bracket is shipped with every patient adjustable
warming unit; the brackets for mounting the patient adjustable
warming unit to an IV pole, bedrail, or a suction wall mount
adapter are available separately. Please contact your local sales
representative or call 3M at 1-800-228-3957 (USA only) for more
information about ordering mounting brackets and adapter.
The following instructions for mounting the patient adjustable
warming unit assume that the brackets have already been attached to
the wall or to the patient adjustable warming unit.
Placing the Patient Adjustable Warming Unit on the Wall-
mount Bracket
1.
Snap the patient adjustable warming unit into position by sliding
the clip on the back of the patient adjustable warming unit into
the groove of the wall-mount bracket. The assembly will click
when the clip is fully seated in the wall-mount bracket.
2.
Lift the patient adjustable warming unit straight up to remove it
from the wall-mount bracket.
Note: It may be necessary to gently strike the case bottom in an
upward fashion with the palm of your hand to release the clip
from the wall-mount bracket.
System
Case
bottom

Mounting the Patient Adjustable Warming Unit on an IV Pole

Caution: If mounted on an IV pole, the distance from the bottom of
the patient adjustable warming unit to the floor must be less than 44"
(112 cm) and the IV pole wheelbase diameter must be at least 28"
(71 cm) to prevent tipping.
1.
To prevent tipping, clamp the Model 875 patient adjustable
warming unit to an IV pole at a height that provides stability.
Clamping the patient adjustable warming unit no higher than
44" (112 cm) from the floor on an IV pole with a minimum 28"
(71 cm) diameter wheelbase.
2.
Turn the clamp knob on the IV pole-mount clockwise to tighten
the clamp to the pole.
en / 34-8726-1742-7
3.

Mounting the Patient Adjustable Warming Unit on a Bedrail

1.
2.
3.
Wall mount
bracket
Clip
Mounting the Patient Adjustable Warming Unit with a Rail-
mount to the Wall
Note: The rail-to-wall bracket is intended for use only with a Bair
Hugger 800 series patient adjustable warming unit.
1.
2.
3.
Snap the patient adjustable warming unit into position by sliding
the clip on the back of the patient adjustable warming unit into
the groove of the IV pole-mount. The assembly will click when
the clip is fully seated in the IV pole-mount.
System
Clamp
knob
IV pole
If the patient adjustable warming unit is not already connected
to the rail-mount, slide the clip on the back of the patient
adjustable warming unit into the groove of the rail-mount. The
assembly will click when the clip is fully seated in the rail-
mount.
Hang the Model 875 patient adjustable warming unit on a
bedrail by the hooks on the rail-mount.
Wrap the safety strap around the bedrail and connect the
end to the fastener on the strap. The safety strap will prevent
the patient adjustable warming unit from falling if the unit is
inadvertently dislodged from the bedrail.
Bedrail
Rail mount
System
If the patient adjustable warming unit is not already connected
to the rail mount, slide the clip on the back of the patient
adjustable warming unit into the groove of the rail-mount. The
assembly will click when the clip is fully seated in the rail-
mount.
Hook the patient adjustable warming unit with the attached rail-
mount to the rail-to-wall bracket.
Lift the patient adjustable warming unit straight up to remove it
from the rail-to-wall bracket.
Clip
IV pole
mount
Safety
strap
5