Notice - 3M Bair Hugger 875 Manuel De L'utilisateur

L'unité de réchauffement réglable par le patient
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
WARNING: To reduce the risk of patient injury or death due to
altered drug delivery:
Do not use a warming gown over transdermal medication
patches.
WARNING. To reduce the risk of injury due to interference with
ventilation:
Do not allow the warming gown or head drape to cover the
patient's head or airway when patient is not mechanically
ventilated.
WARNING: To reduce the risk of strangulation:
Do not leave children and other vulnerable patient populations
unattended while using a Bair Hugger patient adjustable
warming unit.
Do not allow children and other vulnerable patients to handle
the handheld temperature controller without supervision to
prevent controller's electrical cord from becoming entangled
around the patient.
WARNING: To reduce the risks associated with hazardous voltage
and fire:
Do not disassemble the patient adjustable warming unit unless
you are a qualified service technician. There are electrically
live parts within the patient adjustable warming unit when
it is connected to a power source, even when the handheld
controller is in the OFF position.
Keep power cord visible and accessible at all times. The plug on
the power cord serves as the disconnect device.
Only connect to outlets marked "Hospital only," "Hospital
Grade," or a reliable grounded outlet.
Use only the power cord specified for this product and certified
for the country of use.
Do not allow the power cord to get wet.
Do not use the patient adjustable warming unit when it appears
the patient adjustable warming unit, power cord or any
component is damaged. Contact 3M Technical Support.
CAUTION: To reduce the risk of hypothermia and ensure optimal
heat transfer:
Always extend the warming gown insert to its full length when
connected to a warming unit.
Do not tape, roll, gather, bunch, or compress the inflated air-
channel insert of the gown as these actions can occlude airflow
and reduce warming of the patient.
Do not cover gown with more than one sheet.
CAUTION: To reduce the risk of cross-contamination:
This warming gown is not sterile and is intended for single
patient use ONLY. Placing a sheet between the warming gown
and the patient does not prevent contamination of the product.
Clean the patient adjustable warming unit, handheld controller,
and the outside of the warming unit hose after each patient use.
See "Cleaning Instructions" section in this manual.
Follow applicable regulations when disposing of this warming
unit or any of its electrical components.
Do not attempt to clean the air filter as it may be contaminated
from use. Discard the filter in a manner consistent with
institutional protocol.
CAUTION: To reduce the risk of patient or caregiver injury:
Do not use the Bair Hugger pediatric warming gown unless the
patient is over 102 cm (40 inches) tall.
If mounted on an IV pole, the distance from the bottom of the
patient adjustable warming unit to the floor must be less than
4
44" (112 cm) and the IV pole wheelbase diameter must be at
least 28" (71 cm) to prevent tipping.
When using the suction wall mount adapter, only hold the front
of the adapter when increasing/decreasing the expander to
prevent pinching fingers.
CAUTION: To reduce the risk of fire:
This product is classified as Class I Normal Flammability
as defined by the Consumer product Safety Commission's
flammable fabric regulation, 16 CFR 1610. Follow standard
safety protocols when using high intensity heat sources.
CAUTION: To reduce the risk of thermal injury, hyperthermia or
hypothermia:
3M recommends continuously monitoring core temperature. In
the absence of continuous monitoring, monitor the temperature
of patients who are incapable of reacting, communicating
and/or who cannot sense temperature a minimum of every 15
minutes or according to institutional protocol.
Monitor cutaneous responses of patients who are incapable of
reacting, communicating and/or who cannot sense temperature
a minimum of every 15 minutes or according to institutional
protocol.
Adjust air temperature or discontinue warming therapy when
the therapeutic goal is reached, if elevated temperatures are
recorded or if there is an adverse cutaneous response in the
warmed area.
Do not place the patient adjustable warming unit on a soft
uneven surface, such as a bed, or a visibly wet surface as the air
intake may become blocked and cause the patient adjustable
warming unit to overheat, compromising the patient adjustable
warming unit's performance.
Perform all calibrations of the patient adjustable warming unit
using a Model 90055 Temperature Test Kit.
Perform calibration testing at an ambient temperature similar to
the area where the patient adjustable warming unit will be used.
The patient adjustable warming unit is intended for use in an
ambient environment between 18 and 26°C (65 and 79°F).
CAUTION: To reduce the risk associated with electromagnetic
interference (EMI) due to portable and mobile radio frequency (RF)
communications equipment:
The 3M Bair Hugger temperature management system has
been tested to be resistant to both EMI and electrostatic
discharge (ESD).
Install and put into service the 3M Bair Hugger temperature
management system according to the electromagnetic
compatibility (EMC) information provided in the Guidance and
Manufacturer's Declaration.
Should interference occur, move away from the portable or
mobile RF communications equipment.

NOTICE:

1.
The patient adjustable warming unit meets the international
electronic interference requirements. If radio frequency
interference with monitoring equipment occurs, connect the
patient adjustable warming unit to a different power source.
2.
To avoid Bair Hugger patient adjustable warming unit damage:
Do not immerse the patient adjustable warming unit, patient
adjustable warming unit parts, or accessories in any liquid or
subject them to any sterilization process.
Do not use cleaning solutions with greater than 80% alcohol
or solvents, including acetone and thinner, to clean the patient
adjustable warming unit or hose. Solvents may damage the
labels and other plastic parts.
3M™Bair Hugger™ Patient Adjustable Warming Unit Model 875 Operator's Manual

Publicité

Table des Matières
loading

Table des Matières