The 3M™ Bair Hugger™ Patient Adjustable Warming Unit, Model 875 is intended for use in the electromagnetic environment specified
below. The customer or the user of the 3M Bair Hugger Patient Adjustable Warming Unit, Model 875 should assure that it is used in such an
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
Flicker emissions
The 3M Bair Hugger Patient Adjustable Warming Unit, Model 875 is intended for use in the electromagnetic environment specified below.
The customer or the user of the 3M Bair Hugger Patient Adjustable Warming Unit, Model 875 should assure that it is used in such an
Immunity Test
Electrostatic
discharge (ESD)
Electrical fast transient/
burst
Surge
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
en / 34-8726-1742-7
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Compliance
Group 1
Class B
Class A
Complies
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC 60601 Test Level
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line to line
±1 kV line(s) to earth
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
environment.
Electromagnetic Environment - Guidance
The 3M Bair Hugger Patient Adjustable Warming Unit, Model 875 uses RF
energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
The 3M Bair Hugger Patient Adjustable Warming Unit, Model 875 is suitable
for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
environment.
Compliance Level
±8 kV contact
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
±15 kV air
relative humidity should be at least 30%.
±2 kV for power supply
Mains power quality should be that of a typical
lines
commercial or hospital environment.
±1 kV for input/output
lines
±1 kV line to line
Mains power quality should be that of a typical
commercial or hospital environment.
±1 kV line(s) to earth
<5% U
Mains power quality should be that of a typical
T
(>95% dip in U
)
commercial or hospital environment. If the
T
for 0,5 cycle
user of the 3M Bair Hugger Patient Adjustable
Warming Unit, Model 875 requires continued
40% U
operation during power mains interruptions,
T
(60% dip in U
)
it is recommended that the 3M Bair Hugger
T
for 5 cycles
Patient Adjustable Warming Unit, Model 875 be
70% U
powered from an uninterruptible power supply
T
(30% dip in U
)
or a battery.
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
Electromagnetic Environment - Guidance