CAUTION:
1� To reduce the risks associated with instability, impact and facility medical device damage:
•
Mount the Ranger pressure infusor Model 145 only on a 3M model 90068/90124 pressure infusor I�V� pole/base�
•
Do not mount this unit more than 56" (142 cm) from the floor to the base of the pressure infusor unit�
•
Do not use the power cord to transport or move the device�
•
Ensure the power cord is free from the castors during transport of the device�
•
Do not push on surfaces identified with the pushing prohibited symbol�
•
Do not pull the I�V� pole using the pressure infusor power cord�
2� To reduce the risks associated with environmental contamination:
•
Follow applicable regulations when disposing of this device or any of its electronic components�
3� This product is designed for pressure infusion only�
NOTICE:
1� The Ranger pressure infusor meets medical electronic interference requirements� If radio frequency interference with other equipment should occur,
connect the pressure infusor to a different power source�
2� Do not use cleaning solutions with greater than 80% alcohol or solvents, including acetone and thinner, to clean the warming unit or hose� Solvents may
damage the labels and other plastic parts�
3� Do not immerse the Ranger unit or accessories in any liquid or subject them to any sterilization process�
Overview and Operation
Overview and Operation
The pressure infusor system is attached to a custom I�V� pole with base� The power cord is secured allowing the system to be moved around the hospital/healthcare
facility to various locations where needed� The system is designed for frequent use whenever pressure infusion is necessary�
For operation, the Ranger pressure infusor is mounted on the custom I�V� pole and base� A fluid bag is placed in one of the pressure infusor chambers, the infusor's main
power switch is turned ON, and the infusor chambers are activated using the user interface at the front of the device� Upon activation, the air is routed to the pressure
infusor bladder, the bladder begins to inflate and pressurize the fluid bag� The user interface indicates when the fluid bag pressure is In Range and ready for use�
Note: The solution pressure of the Ranger pressure infusor is dependent on surface area and volume of the solution bag� To verify the pressure, refer to the preventive
maintenance manual�
The Ranger pressure infusor has no user adjustable controls� The user slides an I�V� solution bag behind the metal fingers and against the inflation bladder located
inside the pressure infusor� The Ranger pressure infusor should only be used in healthcare facilities by trained medical professionals and can be used within the patient
environment�
When the infusor is attached to an external power source and the main power switch is turned ON, pushing the pressure infusor start/stop button ON inflates the
inflation bladder and maintains pressure on blood and solution bags�
Turning the pressure infusor start/stop button OFF deflates the bladder� Turn the main power switch OFF when the pressure infusor is not in use
3M
Ranger
™
™
Pressure Infusor
High
High
In Range
In Range
300 mmHg
Low
Low
Power entry module
Power entry module
Main power switch
4
Main power switch