Specifications and Technical Data

Parameter
Areas (Channels)
Amplitude
Rate
Width
Cycle
Ramp ON
Contacts
a. Only one Area is available if the rate is >130 pps.
b. Amplitude × Width ≤ 12.7 µC

Materials

Case
Header
Strain Relief
Size/Volume
Production
Boston Scientific Confidential. Unauthorized use is prohibited.
Range
4
0 – 25.5 mA
2 – 1200 pps
a
20 – 1000 μsec
b
1 sec – 90 min, OFF
1 – 10 secs
1 – 16, case: +100% to -100%, OFF
Titanium
Epoxy
Silicone
51.3 mm x 46.0 mm x 10.8 mm / 19.8 cm
Montage MRI IPG Entrada 2.0 19L
Specifications and Technical Data
Default
0 mA
40 pps
210 μsec
OFF
3 secs
1 – 16, case: OFF
(including header)
3
Implantable Pulse Generator
91053242-08 Rev A
92070454 A.1