Table 4: Immunity Testing
RFID Readers
The external electrical components of the Boston Scientific DBS System have been tested for immunity to
interference from RFID readers per the following specifications.
RFID Spec Per AIM 7351731
ISO 14223
ISO/IEC 14443‑3 (Type A)
ISO/IEC 14443‑4 (Type B)
ISOAEC 15693 (ISO 18000‑3 Mode 1)
ISO 18000‑3 Mode 3
ISO/IEC 18000‑7
ISO/IEC 18000‑63 Type C
ISO/IEC 18000‑4 Mode 1
Frequency
134.2 kHz
13.56 MHz
13.56 MHz
13.56 MHz
13.56 MHz
433 MHz
860-960 MHz
2.45 GHz
Vercise™ Deep Brain Stimulation Systems Information for Prescribers
Electromagnetic Compatibility
Test Level (RMS)
65 A/m
7.5 A/m
7.5 A/m
5 A/m
12 A/m
3 V/m
54 V/m
54 V/m