Vercise™ Deep Brain Stimulation Systems Information for Prescribers
Essential Performance
External Trial Stimulator
The stimulation pulse shall meet the requirements for charge balance and amplitude while
stimulation is on.
Other External Devices
Failure of the external electrical components will not result in an unacceptable risk to the user.
Telemetry Information
Vercise PC and Vercise Gevia DBS Systems
The following parameters describe the wireless communication link between the Stimulator and
Remote Control:
•
Frequency Band: 119 kHz to 131 kHz
•
Modulation Type: FSK
•
Effective Radiated Power: 0.05 mW (‑13 dBm) maximum
•
Magnetic Field Strength (at 3 m distance): 46 μA/m
Vercise Genus DBS System
The following parameters describe the wireless communication link between the Remote Control or
Clinician Programmer and the Implantable Pulse Generator or External Trial Stimulator:
•
Frequency Band: 2.402 GHz to 2.480 GHz
•
Modulation Type: GFSK
•
Maximum Radiated Power: 5 dBm
•
Protocol: Bluetooth Low Energy technology