Vercise™ Deep Brain Stimulation Systems Information for Prescribers

Essential Performance

External Trial Stimulator

The stimulation pulse shall meet the requirements for charge balance and amplitude while
stimulation is on.
Other External Devices
Failure of the external electrical components will not result in an unacceptable risk to the user.

Telemetry Information

Vercise PC and Vercise Gevia DBS Systems
The following parameters describe the wireless communication link between the Stimulator and
Remote Control:
Frequency Band: 119 kHz to 131 kHz
Modulation Type: FSK
Effective Radiated Power: 0.05 mW (‑13 dBm) maximum
Magnetic Field Strength (at 3 m distance): 46 μA/m
Vercise Genus DBS System
The following parameters describe the wireless communication link between the Remote Control or
Clinician Programmer and the Implantable Pulse Generator or External Trial Stimulator:
Frequency Band: 2.402 GHz to 2.480 GHz
Modulation Type: GFSK
Maximum Radiated Power: 5 dBm
Protocol: Bluetooth Low Energy technology
Vercise™ Deep Brain Stimulation Systems Information for Prescribers
92328649-04
24 of 149