• A flow of spray from the atomiser indicates that the device is
functioning properly.
6.6 Treatment
• When inhaling, sit upright and relaxed at a table and not in
an armchair to avoid compressing the airways and therefore
impairing the e ectiveness of the treatment.
• Inhale the nebulised medicine deeply.
IMPORTANT
The device is not suitable for continuous operation; after 30 mi-
nutes of operation it must be switched o for 30 minutes.
During treatment, hold the atomiser straight (vertical),
otherwise the atomisation does not work and correct func-
tion cannot be guaranteed.
IMPORTANT
Essential oils, cough medicines, solutions designed for gargling,
and drops for application to the skin or for use in steam baths are
wholly unsuitable for inhalation using a nebuliser. These substan-
ces are often highly viscous and can impair the correct function
of the device and hence a ect the e ectiveness of its application
in the long term.
In the case of oversensitivity of the bronchial system, medicines
containing essential oils can sometimes cause acute bronchio-
spasm (a sudden, cramp-like tightening of the bronchi accompa-
nied by breathlessness). Always ask your doctor or pharmacist
for advice!
6.7 End of inhalation
If the spray is only output sporadically or if the noise changes
during inhalation, you can end the treatment.
• Switch o
the nebuliser after treatment using the on/o
switch [3] and disconnect it from the mains.
• After use, replace the atomiser [9] in its holder [7].
6.8 Nasal douche
The nasal douche is not included in delivery with the IH 21. You
can order the nasal douche using the item number 601.37.
Abdeckung
19
Au angbehälter für Nasensekret
20
Kochsalzbehälter
21
Abdeckloch (Starten der Anwendung)
22
You can use the nasal douche to clean your nasal cavities. The
nasal douche creates a fine jet of liquid for treating respiratory
tract problems or for moistening the nasal mucous membranes..
WARNING
• Do not use the nasal douche in combination with other medi-
cation or essential oils.
IMPORTANT
• The nasal douche should only be used with isotonic saline
solutions.
• Replace the nasal douche after a year.
The maximum volume of the nasal douche is 10 ml. If you are
unsure about the correct dilution of the saline solution, please
consult your doctor.
6.9 Using the nasal douche
To use the nasal douche correctly, proceed as follows
1. Twist the cover [19] 90° anticlockwise and remove it in an up-
ward direction.
2. Lift the collection container [20] out.
3. Fill the saline container [21] with saline solution.
4. Place the collection container [20] into the saline container [21].
5. Place the cover [19] on the saline container [21]. Twist the co-
ver [19] in a clockwise direction until it clicks firmly into place.
6. Connect the compressed air hose [8] to the saline container
[21] from below.
7. Position the cover [19] carefully inside one of your nostrils. Brea-
the through the nose.
8. Press the On/O switch [2] on the back of the nebuliser. To
start the application, place one finger over the cover hole [22]
on the container.
During the application, breathe slowly and evenly
through the nose. Tilt your head slightly in the opposite
direction to the nostril being treated so that the saline
solution can flow deep into the nose.
9. To pause or stop the application, take your finger o the cover
hole [22] on the saline container [21].
10. Once your nose feels clear again, you can stop the applica-
tion. The application should be stopped at the latest once the
maximum filling quantity has been used.
6.10 Clean the device
For information about this, see "Cleaning and disinfection".
7. CHANGING THE FILTER
In normal operating conditions, the air filter must be replaced af-
ter approx. 500 operating hours or one year. Please check the air
filter regularly (after 10-12 nebulisation procedures).
If the filter is very contaminated or blocked, replace the used fil-
ter. If the filter has become wet, it must also be replaced with a
new one.
CAUTION
• Do not try to clean and re-use the used filter!
• Only use original filters supplied by the manufacturer. Other-
wise, your nebuliser may become damaged or e ective the-
rapy can no longer be guaranteed.
• Do not repair or maintain the air filter while it is in use.
• Never operate the device without a filter.
To replace the filter, proceed as follows:
CAUTION
• First switch o the unit and unplug it.
• Allow the unit to cool down.
1. Remove filter cap [5] upwards.
If the filter remains in the unit when the cap is removed,
remove the filter from the unit using tweezers or similar.
2. Replace the filter cap [5] containing the new filter.
3. Check that the filter cap is firmly in place.
15
: