Programming with the
Infinion™ 16 Lead
For detailed programming instructions, refer
to the appropriate Bionic NAVIGATOR™
Software Guide. When the Infinion™ 16
lead is used with the Splitter 2x8 and
placed properly into the Precision™ IPG
ports (laser-marked tail of Splitter 2x8
connecting to IPG port "1-L"), the distal 8
contacts of the Infinion™ 16 lead (contacts
1-8) will correspond with contacts 1-8 on
the left side of the BionicNavigator software
display. Likewise, the proximal 8 contacts
of the Infinion™ 16 lead (contacts 9-16) will
correspond with contact 9-16 on the right side
of the BionicNavigator software display.
Connecting to the IPG
The diagram below depicts this wiring.
Using test stimulation, enlist patient feedback
to verify lead placement and pain coverage.
The user may program with the Infinion™
16 lead using all the functions available in
BionicNavigator for the Linear™ family of
leads (E-Troll, Navigator, etc.). For complete
instructions on programming, see the
BionicNavigator Software Guide.