Precision™ Spinal Cord Stimulator System Clinician Manual
Note: If you experience difficulty when
inserting the lead, lead extension,
or port plug, use the hex wrench
to loosen (counterclockwise) the
setscrew and/or gently rotate the lead
to help advance the proximal end.
Note: To confirm good connections, check
impedances before tightening the
setscrew.
2. Pass the hex wrench through the hole in
the septum located on the top of the IPG
header and tighten each set screw until the
hex wrench "clicks," indicating lock.
CAUTION: Ensure that the lead is fully
inserted before tightening the setscrew to
prevent lead damage.
Note: • If a port plug is used, it is still
necessary to tighten the setscrew on
the port plug, as described above.
• The hex wrench is torque-limited
and cannot be overtightened.
3. Place the IPG in the subcutaneous pocket
with label facing towards the skin.
4. Coil excess lead, extension or splitter un-
der the IPG.
Clinician Manual
90607862-22 Rev A
36 of 303
CAUTION: Do not excessively bend the
Splitter 2x8 receptacle. Bending the re-
ceptacle at a radius tighter than 2" (5 cm)
may result in loss of stimulation via one or
more contacts. Ensure the correct lengths
of leads are selected to prevent excessive
coiling.
5. Secure the IPG in the pocket by suturing
through the holes in the IPG header.
6. Close and dress the wound(s).