Note: On the 2x4 Splitter, the shorter splitter
receptacle corresponds to contacts
1-4, while the longer receptacle
corresponds to contacts 5-8. Make
note of which lead is connected to
each splitter receptacle.
Note: On the Splitter 2x8, one tail is laser-
marked with bands, and corresponds
to contacts 1-8 on the 16-contact
percutaneous lead; the unmarked tail
corresponds to contacts 9-16.
8. If lead repositioning is required, disconnect
the splitter and reinsert the stylet before
advancing the lead. Repeat steps 5 to 7
until satisfactory lead position is achieved.
9. Use the hex wrench supplied to tighten
down the setscrews until the wrench clicks.
Clinician Manual
90607862-22 Rev A
22 of 303
Medical Adhesive
10. If using the Splitter 2x8, prior to closing
the wound, clean the top of the splitter set-
screw seal plug and use silicone medical
adhesive (e.g. Dow Corning Silastic
ical Adhesive Silicone, Type A - Sterile, as
available from Boston Scientific, part num-
ber SC-4320) to coat and seal the top of
the seal plug that has been penetrated by
the hex wrench.
Note: Inadvertent damage to the septum
seal may lead to unintended
stimulation at the splitter 2x8 if the
Medical Adhesive is not used as
intended.
11. Proceed to "Connecting to the Trial Stimu-
lator" on page 32 or "Connecting to the
IPG" on page 35.