Precision™ Spinal Cord Stimulator System Clinician Manual
are disabled in the patient Remote Control.
Access to these parameters can be enabled
through the BionicNavigator software.
To Access the Pulse Width and Rate Options
(if enabled through Bionic Navigator):
1. Press and hold the
"
Control for three seconds until the Pulse
Width screen appears. This is indicated
by "Width" in the bottom left corner of the
screen.
2. Area 1 will initially be highlighted. Press
the
button to cycle to the Area you wish
"
to adjust.
3. When the desired Area is highlighted,
press p or q to increase or decrease the
Pulse Width.
4. To continue to Rate adjustment, press and
hold the
button for three seconds from
"
Clinician Manual
90607862-22 Rev A
button on the Remote
46 of 303
any Pulse Width Area screen. The Rate
screen is indicated by "Rate" in the bottom
left corner of the screen.
5. Area 1 will initially be highlighted. Press
the
button to cycle to the Area you wish
"
to adjust.
6. When the desired area screen is highlight-
ed, press p or q to increase or decrease
the Rate.
7. To return to the main Level screen, press
and hold the
button for three seconds
"
until the Level screen appears.

Device Linking

A Remote Control can communicate with only
one Stimulator at a time. This prevents the
Remote Control from accidentally controlling
an unintended device. In the linking process,
the Remote Control will identify, by telemetry,
the intended Stimulator for communication.
1. Press and hold the
Remote Control and initiate communica-
tion between the Remote Control and the
Stimulator.
2. If the Remote Control is not linked to a
Stimulator, it will display a "Link?" screen.
P
button to unlock the