Precision™ Spinal Cord Stimulator System Clinician Manual
Connecting to the Trial
Stimulator
1. If two leads are used, connect the cable
labeled "1-L" to the upper or left lead, and
the cable labeled "2-R" to the lower or right
lead. Labels are provided. If two 2x4 split-
ters are used, connect the cable labeled
"1-L" to the proximal connector of one of
the splitters and connect the cable "2-R" to
the proximal connector of the other splitter.
If using a Splitter 2x8 connect the cable la-
beled "1-L" to the laser-marked tail and the
cable labeled "2-R" to the unmarked tail.
2. Connect the OR cables to the Trial
Stimulator.
If only one lead is used, connect the OR
cable to "1-L" on the Trial Stimulator.
3. Fit the Trial Belt to the patient, cut off the
excess length, and place the Trial Stimula-
tor in the belt pocket.
Clinician Manual
90607862-22 Rev A
32 of 303
1. OR Cable
2. OR Cable Connector
3. External Trial Stimulator (ETS)
4. Trial Belt
5. Lead or Extension