Boston Scientific Precision Manuel page 45

Masquer les pouces Voir aussi pour Precision:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2. To save/store changes, select the appro-
priate program number and press q. A
confirmation screen will appear.
P
3. Press
to confirm and overwrite, or press
the
button to cancel the operation.
"
Restoring Programs:
The original clinic programs can be restored
by following the instructions below.
P
1. Press and hold
for approximately 3 sec-
onds to access the Restore screen. Pro-
gram 1 will be highlighted.
P
2. Press the
button to cycle through the
programs and select the program to be
restored.
3. Press p to restore the clinic-programmed
settings.
Rechargeable Stimulator System
Coverage Area Selection
Each program can consist of up to 4 different
coverage areas (Areas). The stimulation
strength and other settings for each Area can
be adjusted.
1. Press the
button to access the Area
"
level screen. This controls the stimulation
strength for each Area individually. Press
the
button as needed to highlight a spe-
"
cific stimulation coverage area.
2. Press the p or q button to adjust the
stimulation strength (amplitude) of the se-
lected Area.
Additional Area Options
The Pulse Width and Rate parameters are
available as programming options to allow
patients greater control of their therapy. By
default, access to Pulse Width and Rate
90607862-22 Rev A
Clinician Manual
45 of 303

Publicité

Table des Matières
loading

Table des Matières