SCALE SYSTEM LCD DISPLAY REPLACEMENT (OPTIONAL)

Tools Required:
• Small Regular Screwdriver
• #2 Phillips Screwdriver
Note
Unless otherwise indicated, refer to Figure 5
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Unplug the power cord from the wall receptacle .
3 .
Using a small regular screwdriver, lift and remove the five round self-adhesive screw-covers (B) located on the foot
board cover as well as the self-sticking membrane (A) of the foot board control panel to expose the screws holding
the cover to the foot board . Use caution when inserting the screwdriver under the membrane to avoid scratching
the plastic cover .
Note
Do not reuse the self-adhesive parts due to their adhesive coating considerably losing its effectiveness once they are
removed . Replace them with new parts .
4 .
Properly ground yourself (see the "Static Protection Procedure",
5 .
Using a #2 Phillips screwdriver, remove the seven screws (C) holding the cover to the foot board and remove the
cover after having removed the cables connected to it . Note the cable positions in order to replace them properly .
6 .
Using a small regular screwdriver, push aside the clips (D) and remove the LCD display (E) after having disconnected
its cable .
7 .
Reverse the above steps to install the new LCD display .
8 .
Test the LCD display for proper operation before replacing the foot board cover and returning the bed to service .
Return To Table of Contents
1-34
Service Information
(page
1-33) for this procedure .