FOWLER MOTOR REPLACEMENT

Tools Required:
• Needle Nose Pliers
• #2 Phillips Screwdrivers
• #3 Phillips Screwdrivers
• Pliers
• Cutting Pliers
• 5/16" Socket
• 3/8" Drive Ratchet
• Bungee Cord (or equivalent)
• Small Regular Screwdriver
J
K
E
A
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Lower the Fowler to the flat (0°) position .
3 .
Raise the Gatch to the highest position .
4 .
Unplug the power cord from the wall receptacle .
5 .
Raise the siderails .
6 .
Remove the headboard and footboard .
7 .
Lift and fold the foot section back toward the head end of the bed . Secure it to the bed using a bungee cord .
8 .
Using needle nose pliers, remove the rue ring cotter, washers, clevis pin (A) linking the actuator tube to the seat
section lever arms .