SPEAKER/SIDERAIL INNER PANEL MEMBRANE REPLACEMENT

Tools Required:
• Small Regular Screwdriver
• #2 Phillips Screwdriver
• #3 Phillips Screwdriver
• Cutting Pliers
• Bungee Cord (or equivalent)
• 5/32" Allen Wrench
Procedure:
Note
Unless otherwise indicated, refer to Figure 3
1 .
Follow steps 1 to 4 of the "Siderail Outer Panel Membrane
Replacement" procedure
2 .
Secure the siderail to be repair to the adjacent siderail using a
bungee cord (see Figure 4) .
3 .
Using cutting pliers, cut the lower cable tie (E) holding the
bottom part of the grey cable (there may be two cables) to the
aluminum structure . The slack in the cable will be necessary for
the following steps .
Note
Make sure the cable ties are inserted in the aluminum structure before
reinstalling the structure .
4 .
Using a #2 Phillips screwdriver, remove the screw (F) holding each siderail arm cover and lift up the covers to
remove them .
5 .
Using a 5/16" Allen wrench, remove the two Allen screws (G) holding each pommel to the siderail arms . Leave one
screw loosened on each pommel until ready to remove both pommels .
6 .
While supporting the rail, remove the two last screws and the bungee cord .
7 .
Lift the siderail, pass it over the mechanism arms and lay it on the mattress support while gently pulling on the
cables .
8 .
Using a #3 Phillips screwdriver, remove the six screws (H) holding the aluminum structure to the siderail .
9 .
Grasp and lift the upper part of the aluminum structure until it may be secured temporarily to one of the mechanism
arms using a cable tie .
To replace the speakerphone, proceed with steps 10 to 13.
To replace the inner panel membrane, proceed to steps 14 to 17.
10 . Remove the cables connected to the speakerphone . Note their position before removing them .
11 . Using a #2 Phillips screwdriver, remove the four screws (I) holding the speakerphone to the siderail and remove it .
12 . Reverse the preceding steps to install the new speakerphone .
Note
Before reassembling the siderail, it is recommended the continuity is checked at the speakerphone terminals to ensure
the circuit operates properly .
13 . Test the nurse call and all controls of the inner and outer control panels for proper operation before reassembling
the siderail and returning the bed to service .
14 . Disconnect the inner membrane connector .
www.stryker.com
Service Information
(page
1-30) for this procedure .