FOOT END LIFT MOTOR REPLACEMENT (CONTINUED)
14 . Reverse the above steps to install the new motor .
15 . Plug the bed into the wall receptacle .
16 . Lower the bed, and hold the bed down button until the lift motor reaches its lowest position .
CAUTION
The limits of the new motor must be adjusted prior to reconnecting it to the lift lever . An improper adjustment can damage
the lift mechanism .
17 . To adjust the limits of the new motor, proceed as follows:
A . Ensure the motor cable is connected to the control board and plug the power cord into the wall receptacle .
B . Press the bed down control until the motor stops . This will be the lower limit of the motor .
C . Attach the molded nut support on the motor to the lift lever arms .
D . With the alignment fixtures still in place, raise and lower the bed completely to ensure the lower limit is
properly adjusted .
18 . Test the Bed Up and Bed Down functions for proper operation before returning the bed to service .