CONTROL BOARD REPLACEMENT
Tools Required:
• #2 Phillips Screwdriver
• #3 Phillips Screwdriver
• Needle Nose Pliers
• Cutting Pliers
• 3/8" Combination Wrench
A
C
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Unplug the bed from the wall receptacle .
3 .
Remove the foot board .
4 .
Using a #3 Phillips screwdriver, remove the six screws (A) holding the plastic cover to the foot end casing . Grasp
both ends of the cover, move them apart to disengage the inside clips from the round accessory brackets and
remove the cover .
5 .
Using a #2 Phillips screwdriver, remove the four screws (B) holding the protective plate to the foot end casing .
6 .
Properly ground yourself (see "Grounding Diagram"
7 .
Using cutting pliers, cut the cable ties holding together the cables .
Note
• Carefully note the position and the grouping of the cables before clipping the cable ties .
• Ensure Dip Switch settings of the new control board match the ones on the old board .
8 .
Remove the jumper from W1 on the control board and install it on the new board .
9 .
Remove all cables connected to the control board . Note their location so they will be connected properly to the new
board . Refer to drawing OL280120,
on the motor control board .
10 . Using a 3/8" combination wrench, remove the nut/screw (C) holding the ground cable to the frame .
11 . Using needle nose pliers, pinch the upper part of the stand-off pins and lift the defective board up and out .
12 . Using a 3/8" wrench and a #2 Phillips screwdriver, remove the nut/screw (D) holding the ground cables to the board .
13 . Reverse the above steps to install the new control board .
14 . Test all the bed functions before returning the bed to service .
www.stryker.com
Service Information
B
page
page
1-88, or OL280121,
280C-009-002 REV B
D
1-13) .
page
1-94, for the connecting position of the cables
C
Figure 11
Return To Table of Contents
1-47