Lubrication Requirements - Stryker GoBed II MedSurg Bed Guide D'entretien

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

LUBRICATION REQUIREMENTS

The only components of the bed needing periodic lubrication are the four actuator screws and the clevis pins holding the
head actuator to the head section lever . These components should be inspected annually and lubricated every two years .
CAUTION
The bed uses oil impregnated shoulder spacers at hinge points . DO NOT lubricate these shoulder spacers . When
shoulder spacers become worn, replace them .
ACTUATOR SCREW LUBRICATION
Tools Required:
• #2 Phillips Screwdriver
• Brush
• Bungee Cord (or equivalent)
• OG2 Grease (or equivalent)
Procedure:
Clevis Pin
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Remove the plastic head section cover (if equipped) . Raise the head section to the highest position .
3 .
Remove the headboard .
4 .
Unplug the power cord from the wall receptacle .
5 .
Apply a coat of OG2 grease to the clevis pin (A) .
6 .
Raise and lower the head section several times to ensure the clevis pin is properly coated .
Procedure:
Head and Thigh Actuators
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Remove the plastic head and foot section covers (if equipped) .
3 .
Raise the head section to the highest position and flatten the thigh section . Lower the four siderails .
4 .
Unplug the power cord from the wall receptacle .
5 .
Lift the foot section and fold it back towards the head end of the bed . Secure it using a bungee cord .
6 .
Using a #2 Phillips screwdriver, remove the four screws (B) holding the cover plate to the frame and remove the
plate . Remove the night light if equipped .
7 .
Using a #2 Phillips screwdriver, remove the two screws (C) holding the dust tube of each actuator . Push the tube
back to uncover the screw threads .
8 .
Using a brush, apply OG2 grease to the screw threads . Ensure the grease penetrates to the bottom of the threads .
9 .
Replace the dust tubes and cover plate, and lower the foot section .
10 . Plug the power cord into the wall receptacle .
11 . Using the electronic controls, raise and lower the head and thigh section several times to spread the grease evenly .