POWER CORD CONNECTOR REPLACEMENT

Tools Required:
• #1 Phillips Screwdriver
• #2 Phillips Screwdriver
Procedure:
Note
Unless otherwise indicated, refer to Figure 12,
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Unplug the bed power cord from the wall outlet .
3 .
Remove the power cord from the power connector and disengage it from the wire clip .
4 .
Properly ground yourself (see "Grounding Diagram"
5 .
Using a #2 Phillips screwdriver, remove the six screws (A) holding the cover to the litter . Keep the cable clips .
6 .
Once the screws are removed, pivot the cover and lay it flat on the bottom of the litter (see opposite illustration) .
7 .
Remove all cables connected to the power cord connector . Note the location of the cables in order to properly
reconnect them later . Refer to drawing OL280120,
of the cables on the power connector .
8 .
Turn the cover to a vertical position . Using a #1 Phillips screwdriver, remove the two screws (D) holding the power
connector to the cover .
9 .
Unplug the connector and remove it .
Note
Ensure the proper size fuses are used in the new power cord connector before installing it .
10 . Reverse the above steps to install the new power cord connector .
11 . Test all the bed functions before returning the bed to service .
www.stryker.com
Service Information
page
1-48,for reference for this procedure .