Stryker GoBed II MedSurg Bed Guide D'entretien page 58

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

SCALE CALIBRATION (OPTIONAL) (CONTINUED)
3 .
Select CALIB . MENU using the Down key and press ENTER . The display will read:
4 .
Put the inclinometer on the seat section of the mattress support and position the bed to 0° . Then press ENTER .
Display will read:
5 .
Do not touch the bed until the end of this step . When completed, the display will read:
6 .
Put the bed in the Trendelenburg position (head down, foot up) until the inclinometer reads +12° . Press ENTER . The
display will read:
7 .
Do not touch the bed until the end of this step . When completed, the display will read:
8 .
Put the bed in the Reverse Trendelenburg position (head up, foot down) until the inclinometer reads -12° . Press
ENTER . The display will read:
9 .
Once the last step is completed, the system will automatically return to the main maintenance menu .
10 . To return to the normal scale display, press ZERO .
11 . Before returning the bed to service, test the scale accuracy . See the "Verifying Scale Accuracy",
Note
• It is possible to end the calibration process any time before it is completed by pressing the ZERO key . The preceding
calibration data will be restored .
• If the bed is in the incorrect angle in step two or three, the same message will reappear on the display until the bed
is placed at the angle requested on the display .
Return To Table of Contents
1-56
Service Information
PUT BED AT 0°
PRESS ENTER
DO NOT TOUCH BED
PUT BED AT +12°
PRESS ENTER
DO NOT TOUCH BED
PUT BED AT -12°
PRESS ENTER
DO NOT TOUCH BED
280C-009-002 REV B
page 1-55 .
www.stryker.com