SEAT SECTION REPLACEMENT

Tools Required:
• 1/2" Long Socket
• (2) 1/2" Combination Wrenches
• 3/8" Drive Ratchet
• #2 Phillips Screwdriver
Procedure:
Note
Unless otherwise indicated, refer to Figure 6
1 .
Raise the bed to the highest position and apply the brakes . Flatten the mattress support .
2 .
Unplug the power cord from the wall receptacle .
3 .
Lower the siderails .
4 .
Using two 1/2" combination wrenches, remove the locknut, shoulder spacers and bolt holding the mattress support
lever of the CPR mechanism (N, Figure 17,
5 .
Using a 1/2" long socket and a 1/2" combination wrench, remove the four locknuts, four washers and four bolts (E)
holding the seat section to the frame .
6 .
Using a 1/2" combination wrench, remove the four locknuts, four washers, four shoulder spacers and four bolts (C,
F) linking the seat section to the thigh and head section and remove the defective seat section .
7 .
Using a #2 Phillips screwdriver, remove the three screws (G) holding the protective plate to the seat section .
8 .
Reverse the above steps to install the new seat section .
9 .
Test the bed for proper operation before returning the bed to service .
www.stryker.com
Service Information
(page
1-35) for this procedure .
page
1-59) to the seat section .
280C-009-002 REV B