FOWLER REPLACEMENT

Tools Required:
• 1/2" Socket
• (2) 1/2" Combination Wrenches
• #2 Phillips Screwdriver
• 3/16" Allen Wrench
• 7/16" Combination Wrench
• Cutting Pliers
• Strap (or equivalent)
• 3/8" Drive Ratchet
Procedure:
Note
Unless otherwise indicated, refer to Figure 6,
WARNING
Do not activate the CPR emergency release handles during this procedure or serious injury to the maintenance technician
or damage to equipment could occur .
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Unplug the power cord from the wall receptacle .
3 .
Raise the head siderails .
4 .
If applicable, remove the litter plastic cover (if equipped) or raise the fowler to the highest position . Secure the head
section to the bed using a strap to prevent the head section from moving during the procedure .
5 .
Loosen the lock rings (A, Figure 1,
pliers, cut the cable ties holding the siderail cables to the frame .
6 .
Using a 3/16" Allen wrench and a 1/2" socket, remove the four locknuts and four bolts (B, Figure 1,
holding each head siderail assembly to the head section and remove the siderails .
7 .
Using a #2 Phillips screwdriver, remove the screw holding each cable tie present on both sides of the fowler structure .
8 .
Using a 1/2" combination wrench, remove the two nuts, two shoulder spacers and two bolts (H) holding the fowler
to the stabilizer arms . Lay the stabilizer arms down .
If the bed is equipped with the CPR mechanism, proceed with steps 9 to 16.
If the bed is not equipped with the CPR mechanism, proceed to steps 17 to 21.
9 .
Remove the two wires (C, Figure 17,
10 . Using two 1/2" combination wrenches, remove the locknut, shoulder spacer and bolt (L, Figure 17,
holding each CPR handle to the head section . To remove the cable ends from the handle slots, loosen the two
locknuts holding each CPR cable to its handle using a 7/16" combination wrench .
Note
Insert the cable ends inside the handle slots before assembling the handle to the new head section . Then adjust the two
locknuts of each cable in order to have a 1/8" play in the CPR handle movement .
Return To Table of Contents
1-38
Service Information
page
1-35, for this procedure .
page
1-28) of both siderail cables and remove the siderail cables . Using cutting
page
1-59) from the micro switch . Note their position before removing them .
280C-009-002 REV B
page
1-28)
page
1-59)
www.stryker.com