ACTUATOR SCREW LUBRICATION (CONTINUED)
Procedure:
Hi-Lo Actuators
1 .
Lower the bed to the lowest position and apply the brakes . Flatten the litter .
2 .
Lower the four siderails .
3 .
Using a #2 Phillips screwdriver, remove the head and foot section plastic covers (if equipped), to reach the Hi-Lo
actuators, or raise the head section to the highest position and fold the foot section back towards the head end of
the bed . Secure the foot section using a bungee cord .
4 .
Unplug the power cord from the wall receptacle .
5 .
Using a brush, apply OG2 grease to the screw threads of both actuators through the openings on the sides of the
dust tubes . Ensure the grease penetrates to the bottom of the threads .
6 .
Lower the foot section .
7 .
Plug the power cord into the wall receptacle .
8 .
Using the electronic controls, raise and lower the bed several times to spread the grease evenly .
www.stryker.com
Preventative Maintenance
280C-009-002 REV B
Return To Table of Contents
1-17