TRANSFORMER REPLACEMENT (100/200/220/240 VAC OPTIONS)
Tools Required:
• #2 Phillips Screwdriver
• 1/2" Combination Wrench
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Unplug the power cord from the wall receptacle .
3 .
Using a #2 Phillips screwdriver, remove the six screws (A) holding the cover to the litter . Keep the cable clips . See
Figure 12,
page
1-48 .
4 .
Properly ground yourself (see "Grounding Diagram" section,
5 .
Using a #2 Phillips screwdriver, remove the nut, washer and bolt (A) holding the transformer to the litter . Disconnect
the transformer cable and remove it .
Note
Use caution to avoid overtightening the bolt .
6 .
Reverse the above steps to install the new transformer .
7 .
Test the bed for proper operation before returning the bed to service .
Return To Table of Contents