HEAD END SIDERAIL ASSEMBLY REPLACEMENT
Tools Required:
• 1/2" Socket
• 3/8" Drive Ratchet
• 3/16" Allen Wrench
• Cutting Pliers
B
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
If applicable, remove the plastic litter cover (if equipped) . Raise the fowler to the highest position and raise the
siderail to be repaired .
3 .
Unplug the power cord from the wall receptacle .
4 .
Loosen the lock ring (A) and unplug the siderail cable .
5 .
Using cutting pliers, remove the cable ties holding the siderail cable to the frame .
6 .
Using a 3/16" Allen wrench and a 1/2" socket, remove the four locknuts/bolts (B) holding the siderail assembly to
the head section and remove the assembly . Support the assembly when removing the last bolts .
7 .
Reverse the above steps to install the new siderail assembly .
8 .
Test the siderail movement and all controls of both control panels, including the nurse call (optional) and the
communications package (optional), for proper operation before returning the bed to service .
Return To Table of Contents
1-28
Service Information
B
280C-009-002 REV B
A
Figure 1
www.stryker.com