Problem / Failure
No power to bed .
No bed up or down motion .
No Fowler up or down motion .
Fowler does not reach full height .
No Gatch up or down motion .
No Auto Contour motion .
Return To Table of Contents
1-18
Troubleshooting Guide
• Verify power at wall receptacle .
• Verify the power cord connections at the wall receptacle and
the bed .
• Check the power cord for damage . Replace if necessary .
• Check the two fuses inside the power connector .
• Ensure Total Lockout (padlock LED) is not enabled .
• Check the siderail control panel cable connection to the bed
receptacle (under the mattress support) .
• Verify the bed has power .
• Ensure Total Lockout (padlock LED) is not enabled .
• Ensure Fowler Lockout (padlock LED) is not enabled .
• Check the siderail control panel cable connection to the bed
receptacle (under the mattress support) .
• Verify the bed has power .
This situation happens when the CPR handle is used to partly
lower the Fowler . The use of the CPR mechanism for this purpose
creates a situation where the Fowler motor is out of sync with the
actual position of the Fowler . To correct the situation:
• Completely lower the Fowler using the CPR handle or the
Fowler down control to enable the Fowler motor to reset itself .
Refer to the "Emergency CPR Release" section of the Operations
Manual for more information .
• Ensure Total Lockout (padlock LED) is not enabled .
• Ensure Gatch Lockout (padlock LED) is not enabled .
• Check the siderail control panel cable connection to the bed
receptacle (under the mattress support) .
• Verify the bed has power .
• Ensure Gatch Lockout (padlock LED) is not enabled .
• Check the two Auto Contour limit switches . Replace if needed .
280C-009-002 REV B