AUTO CONTOUR MICRO SWITCH REPLACEMENT

Tools Required:
• Bungee Cord (or equivalent)
• #1 Phillips Screwdriver
A
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Raise the thigh section . Remove the plastic litter cover (if equipped) of the fowler section . Lift and fold the foot
section back toward the head end of the bed . Secure it using a bungee cord .
1ST AUTO CONTOUR MICRO SWITCH
This switch tells the control board that the Gatch has reached the preset angle for the Auto Contour position .
3 .
Remove the cables from the switch . Note the cable locations
in order to properly reconnect them to the new switch.
4 .
Using the Gatch down control, slowly lower the thigh section
until only light contact remains between the activator (A)
attached to the knee section and the push-button switch (B) .
The replacement procedure will start from this position (see
opposite illustration) .
5 .
Unplug the bed power cord from the wall outlet .
6 .
Using a #1 Phillips screwdriver, remove the two screws (C)
holding the support to the frame .
7 .
Press the two clips on each side of the switch to remove it from
the support .
8 .
Reverse the above steps to install the new micro switch .
9 .
Test the Auto Contour positioning for proper operation before
returning the bed to service .
www.stryker.com
Service Information
Figure 16
280C-009-002 REV B
E
D
B
C
Return To Table of Contents
1-57